1: Unpack and Inspect the Synergy H1
The Synergy H1 with all available modules weighs up to 55 pounds
(24.95 kg). Use two people when lifting and carrying the instrument.
Save all packaging materials. If you need to ship the reader to BioTek for
repair or replacement, you must use the original materials. Using other
forms of commercially available packaging, or failing to follow the
repackaging instructions, may
void the warranty
. Improper packaging
the results in damage to the reader may lead to additional charges.
During the unpacking process, inspect the packaging, reader, and
accessories for shipping damage. If the reader is damaged, notify the
carrier and your BioTek representative. Keep the shipping boxes and the
packaging materials for the carrier’s inspection. BioTek will arrange for
repair or replacement of your reader immediately.
1. Open the shipping box. Remove the instrument and place it on a level, stable surface.
2. Place the packaging materials back into the shipping box for reuse if the instrument
needs to be shipped again.
2: Select an Appropriate Location
Install the reader on a level, stable surface in an area where ambient temperatures
between 18°C (64°F) and 40°C (104°F) can be maintained. Leave at least six inches of
space between the instrument’s rear panel and any other object. This space ensures proper
air flow in and out of the instrument.
The reader is sensitive to extreme environmental conditions. Avoid the following:
l
Excessive humidity.
Condensation directly on the sensitive electronic circuits can
cause the instrument to fail internal self-checks. The humidity must be in the range
of 10–85%, non-condensing.
l
Excessive ambient light.
Bright light may affect the reader’s optics and readings,
reducing its linear range.
l
Dust.
Readings may be affected by extraneous particles (such as dust) in the
microplate wells. A clean work area is necessary to ensure accurate readings.
If you will be installing the BioStack for operation with the Synergy H1, you may wish to
seat the instruments in their aligning plates now. Refer to the
BioStack Operator’s
Manual
for more information.
Synergy H1 Operator's Manual
1: Unpack and Inspect the Synergy H1 | 9
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...