Absorbance Liquid Tests
starting on page
describes the test methods, lists the Gen5
protocol parameters, explains how to analyze the test results, and provides
troubleshooting tips in the event of test failure.
Absorbance Liquid Test 1
The tests in this section require specific microplates, solutions, and
filters or wavelengths. Your laboratory may require a deviation from
some of these tests. For example, you may wish to use a different plate
or test solution. If deviation from the tests as presented in this section is
required, perform the following steps the first time each test is run:
Perform the tests exactly as described here.
Rerun the tests using your particular plates, solutions, and so on.
If results are comparable, then the results from these tests will be
your baseline for future tests. Document your new test procedure
and save all test results.
Materials
Manufacturer part numbers are subject to change.
l
New 96-well clear flat-bottom microplate (Corning Costar #3590 recommended)
l
Stock Solution A or B, which may be formulated by diluting a dye solution available
from BioTek (A) or from the materials listed below (B)
l
Gen5 protocol
Synergy H1 Abs Test 1.prt
described on page
Solution A
l
BioTek QC Check Solution No. 1 (PN 7120779, 25 mL; PN 7120782, 125 mL)
l
Deionized water
l
5-mL Class A volumetric pipette
l
100-mL volumetric flask
1. Pipette a 5-mL aliquot of BioTek QC Check Solution No. 1 into a 100-mL volumetric
flask.
2. Add 95 mL of DI water; cap and shake well. The solution should measure
approximately 2.000 OD when using 200 μL in a flat-bottom microwell.
114 | Chapter 8: Instrument Qualification Procedures
BioTek Instruments, Inc.
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...