Parameter
Setting
READ STEP 2
Detection Method
Luminescence
Read Type
Endpoint
Optics Type
Filters
Step Label
Background
Read wells
D1–G4
Filter Set
1 (filter cube)
Excitation
Plug
Emission
Hole
Gain
135
Integration Time
0:10.00 MM:SS.ss
Delay After Plate Movement
350 msec
Dynamic Range
Standard
Read Height
7.00 mm
READ STEP 3
Detection Method
Luminescence
Read Type
Endpoint
Optics Type
Filters
Step Label
Battery check
Read wells
A7–A8
Filter Set
1 (filter cube)
Excitation
Plug
Emission
Hole
Gain
60
Integration Time
0:01.00 MM:SS.ss
Delay After Plate Movement
350 msec
Dynamic Range
Extended
Read Height
10.00 mm
84 | Chapter 7: Instrument Qualification Process
BioTek Instruments, Inc.