These instructions briefly describe how to create an experiment and then read a plate in
Gen5. See the Gen5 Help system for complete instructions.
1. In the Gen5 Task Manager, select the Experiments icon and click
Create using an
existing protocol
.
2. Select the desired protocol and click
OK
.
3. Select a plate in the menu tree and select
Plate > Read Plate #
or click the
Read
New
icon.
4. When the read is complete, measurement values appear in Gen5.
5. Select
File > Save
and give the experiment an identifying name.
Dispense Module Control
Applies only to models equipped with injectors
Gen5 is used to perform several dispense functions, such as initialize, dispense, prime, and
purge. The Prime and Purge functions are introduced here; refer to the Gen5 Help system
for additional information.
Prime
Before running an experiment with a Dispense step, prime the system with the fluid to
be used.
1. Place the priming plate on the carrier.
2. Fill the supply bottle with a sufficient volume of the fluid to be used for the prime
and the assay. Insert the appropriate inlet tube into the bottle.
3. Select
System > Instrument Control > Synergy H1
and click the
Prime
tab.
4. Select the Dispenser number (1 or 2) associated with the supply bottle.
5. Enter the Volume to be used for the prime. The minimum recommended prime
volume is 2000 μL.
6. Select a prime Rate, in μL/second.
7. Click
Prime
to start the process. When finished, carefully remove the priming
plate from the carrier and empty it.
If the priming plate is empty, the prime volume was too low.
Purge
To conserve reagent, Gen5 provides the option to purge fluid from the system back into
the supply bottle.
1. Select
System > Instrument Control > Synergy H1
and click the
Prime
tab.
2. Select the Dispenser number (1 or 2) associated with the supply bottle.
3. Enter the desired purge Volume in μL (e.g., 2000).
Synergy H1 Operator's Manual
Gen5 Software | 37
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...