9: Install Gen5 Software
The Synergy H1 is controlled by Gen5 software running on a host
computer. There is a certain sequence of events that must be followed to
ensure that the software is properly installed and configured. Please
follow the instructions provided in the
Gen5 Getting Started Guide
to
install the software.
10: Turn on the Reader
1. If Gen5 is open, close it now.
2. The power switch is located on the lower-left corner of the front panel; turn on the
Synergy H1. The reader performs a System Test. When the test is completed, the
reader extends the microplate carrier.
The carrier eject button, located next to the reader’s power switch, can be used to
extend/retract the microplate carrier.
11: Establish Communication
If using the USB cable, refer to the instructions that shipped with the USB drivers on the
Gen5 software media to install the necessary drivers.
1. Start Gen5 and log in if prompted.
2. From the main screen select
System > Instrument Configuration
.
3. Click
Add Reader
and select
Synergy H1
. Click
OK
.
4. Perform one of the following steps, as applicable:
l
Select
Plug & Play
. (A reader must be connected to the computer and turned on
to appear in the Available Plug & Play Readers list.)
l
Select
Com Port
and select the computer’s COM port to which the reader is
connected. (If using the USB cable, the information can be found via the Windows
Control Panel, under Ports in the Hardware/Device Manager area of System
Properties.)
5. Click
Test Comm
. Gen5 attempts to communicate with the reader. If the
communication attempt is successful, return to Gen5’s main screen.
If the communication attempt is not successful, try the following:
Synergy H1 Operator's Manual
9: Install Gen5 Software | 17
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...