Repackaging and Shipping Instructions
If the equipment has been exposed to potentially hazardous material,
decontaminate it to minimize the risk to all who come in contact with the
reader during shipping, handling, and servicing. Decontamination prior
to shipping is required by the U.S. Department of Transportation
regulations. See page
for decontamination instructions for the reader
and dispense module.
Remove the microplate and tip prime trough (if equipped) from the
carrier before shipment. Spilled fluids can contaminate the optics and
damage the instrument.
The Synergy H1 with all available modules weighs up to 55 pounds
(24.95 kg). Use two people when lifting and carrying the instrument.
The instrument’s packaging design is subject to change. If the
instructions in this section do not apply to the packaging materials you
are using, please contact BioTek’s Technical Assistance Center.
Replace the shipping hardware before repackaging the reader. Please
contact BioTek and order part number 8040015 if you need a carrier
shipping bracket and/or filter reader shipping bracket.
When preparing to ship the Synergy H1 and/or the dispense module to
BioTek, be sure to use the original packaging materials. Other forms of
commercially available packaging are not recommended and can
void
the warranty
.
The shipping materials are designed to be used no more than five times.
If the original materials have been damaged, lost, or used more than
five times, contact BioTek to order replacements.
Refer to the
Gas Controller User Guide
for the decontamination procedure and packing
instructions for the gas controller module.
1. Contact BioTek’s Technical Assistance Center for a Service Call Notice (SCN) number
and the shipping address.
2. Decontaminate the reader and, if attached, the dispense module, according to the
instructions provided in
starting on page
3. If you will also be shipping the dispense module, perform the steps described on
page
If you are not shipping the dispense module, disconnect it from the reader now.
22 | Chapter 2: Installation
BioTek Instruments, Inc.
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...