Filter-Based Fluorescence
Fluorescence Intensity
Sodium Fluorescein in phosphate buffered saline (PBS)
DL ≤10 pM, top read
Excitation 485/20 nm, Emission 528/20 nm, 510 nm mirror
Methylumbelliferone (MUB) in carbonate-bicarbonate buffer (CBB)
DL ≤0.16 ng/mL, top read
Excitation 360/40 nm, Emission 460/40 nm, 400 nm mirror
Time-Resolved Fluorescence
Europium
DL ≤ 250 fM, top read
Excitation 360/40 nm, Emission 620/40 nm, 400 nm mirror
Integration Time 20 to 16,000 µs, Delay 0 to 16,000 µs, Granularity 1-µs steps
Fluorescence Polarization
Sodium Fluorescein
5 mP standard deviation at 1 nM sodium fluorescein
Excitation 485/20 nm, Emission 528/20 nm, 510 nm mirror
Excitation range 330–700 nm (UV-transparent polarizing filter)
Emission range 400–700 nm
Luminescence Specifications
≤ 75 amol/well DL ATP in a 96-well plate (low-noise PMT), 30 amol typical
≤ 500 amol/well DL ATP in a 96-well plate (red-shifted PMT)
Synergy H1 Operator's Manual
Luminescence Specifications | 143
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...