To remove a syringe: In Gen5, click
System > Instrument Control >
Synergy H1
. On the Prime tab, select a dispenser and click
Maintenance
.
The syringe bracket will move to its furthest-from-home position. Remove
the metal thumbscrew from underneath the bracket. Unscrew the top of the
syringe from the bottom of the syringe drive. Gently remove the syringe.
7. Moisten a cloth with DI or distilled water and wipe all surfaces that have been
cleaned with the bleach solution or alcohol.
8. Rinse all tubing and the syringes with DI water.
9. Use a clean, dry cloth to dry all surfaces on the instrument and the dispense
module.
10. Reassemble the dispense module as necessary.
11. Discard the used gloves and cloths using a biohazard trash bag and an approved
biohazard container.
Dispense Module, Syringe Replacement
Refer to the
Preventive Maintenance
chapter for cleaning procedures you must perform
regularly and also in the case of poor performance (for example, when the Dispense
Accuracy and Precision tests fail). If cleaning the injection system does not eliminate
performance problems, or if a syringe is leaking, perform these instructions to replace a
faulty syringe. Contact BioTek TAC to order replacement syringes.
To change a syringe, first use Gen5 to put the syringe in its maintenance position.
Syringe Maintenance Position
Do not change the syringe position or calibrate the dispensers unless
instructed to do so by BioTek as part of installation, upgrade, or
maintenance.
Gen5 provides access to syringe setup functions for maintenance and calibration purposes.
When a syringe needs to be installed or replaced, it must first be moved to its
“maintenance position.”
1. In Gen5, select
System > Instrument Control > Synergy H1
and click the
Prime
tab (or
Dispenser
tab, if using Gen5 v2.05 or lower).
2. Select the appropriate Dispenser number (1 or 2) associated with the syringe.
3. Click
Maintenance
. The syringe plunger will move to its furthest-from-home
position. The syringe can then be disconnected from the drive bracket and unscrewed
from the valve.
Synergy H1 Operator's Manual
Dispense Module, Syringe Replacement | 71
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...