l
Is the reader connected to the power supply and turned on?
l
Is the communication cable firmly attached to both the reader and the computer?
l
Did you select the correct Reader Type in Gen5?
l
Try a different Com port.
l
Did you install the USB driver software?
l
If you remain unable to get Gen5 and the reader to communicate with each other,
contact BioTek’s Technical Assistance Center.
12: Verify/Set Dispenser Calibration Values
Applies only to models equipped with injectors
Confirm that the reader is configured with calibration values for the dispense module.
i
The calibration values for both dispensers (#1 and #2) are printed on labels affixed to
the rear of the dispense module. Each label lists six target calibration values (e.g., 200,
80, 40) with their actual measured values (e.g., 199.3, 79.7, 39.9). Gen5 should display
the
measured
calibration values.
1. If you have not already done so, turn on the instrument and establish communication
with Gen5.
2. In Gen5, go to
System > Instrument Configuration
, select the
Synergy H1
, and
click
View/Modify
.
3. Click
Setup
and select the
Dispenser 1
tab.
4. Click
Get Volumes
.
5. Compare the Calibration Volumes in the dialog with the Syringe #1 values on the
rear panel of the dispense module.
If the values match, skip to step 6.
If there is a mismatch:
l
Press CTRL+SHIFT+M to enter maintenance mode for the Dispenser 1 window.
l
Enter the syringe calibration values from the corresponding label on the rear of
the dispense module.
l
Click
Send Volumes
and then
Get Volumes
to verify that the entered values
were sent to the reader.
6. Select the
Dispenser 2
tab and repeat steps 4–5 for Dispenser 2.
18 | Chapter 2: Installation
BioTek Instruments, Inc.
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...