If an Absorbance Test Plate is not available,
Absorbance Liquid Test 2
may be
conducted to test the instrument's alignment, repeatability, and accuracy by preparing a
series of solutions of varying OD values as described on page
Absorbance Liquid Test 3
is an optional test offered for those sites that must have
proof of linearity at 340 nm. (Alternatively, the BioTek 340 nm Absorbance Test Plate
may be used; see page
.) This test is optional since the Synergy H1 has good “front-
end” linearity throughout the specified wavelength range. While the absolute values of
the OD cannot be determined by this test, the results will indicate if there is adequate
repeatable absorbance and a linear slope. This method is dependent upon proper dye
dilution and a skilled pipetting technique. It is expected that the first dilution (mid-level
solution) will have an absorbance value near 75% of that of the stock (high-level)
solution, and that the second dilution (low-level solution) will have an absorbance value
near 50% of that of the stock solution.
Gen5 Protocol Parameters
The information in this section represents the recommended reading parameters for the
referenced Gen5 protocol(s). It is possible that your tests will require modifications to
some of these parameters, such as the Plate Type.
i
The Plate Type setting in each Gen5 protocol should match the actual plate in use.
Synergy H1 Abs Test 1.prt
Parameter
Setting
Plate Type
96 WELL PLATE
Shake Step
Linear, 4 minutes, default frequency
Two Read Steps
Kinetic loop (one per Read step)
Set a Run Time/Interval combination to read the
plate five times with minimal delay
Detection Method
Absorbance
Read Type
Endpoint
Optics Type
Monochromators
Read wells
First Read step: A1..H2
Second Read step: A11..H12
Wavelength
405 nm
Read Speed
Normal
Delay after plate movement
100 msec
Plate Out,In step between loops
Text "rotate the plate 180 degrees"
78 | Chapter 7: Instrument Qualification Process
BioTek Instruments, Inc.
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...