VALENT Cart Specifications
Bulk reagent carboys
Six carboys, 8L capacity per carboy.
Secondary containment capacity
11L maximum capacity
Electrical requirements
Power output: 3600W(3600VA)
Power source: 100-120 VAC 50/60 Hz
Requires a step-down transformer rated for 2kW for operation using 220V mains (not
supplied)� Cart must be factory-configured for 50Hz operation�
Do not plug anything into the VALENT Cart besides the computer and label printers�
VALENT Cart dimensions
40”W X 24”D X 30”H / 102cm W X 61cm D X 77cm H
Cart weight
250lbs / 113.4 kg
Undervoltage and
overvoltage protection
The VALENT Cart is protected from undervoltage/overvoltage conditions by internal relays
built into the inverter/charger which will limit input voltage to 85-140VAC. Operation is
prevented if input voltage is outside the specified range.
NOTE: All cart circuit breakers must be in the OFF (down) position before disconnecting
the power cord from the cart or the wall outlet� Failure to do so may cause the batteries
to run down and necessitate a service call�
Reagent Vial Label and Slide Label Printers
The VALENT Software is compatible with Microsoft Windows 10 and will have all computer system software pre-installed. The system
also includes two TSC TC310 Thermal Transfer Printers. One printer is required for slide labels and the second printer is required for
reagent vial labels.
Report Printer
Please see manufacturer’s specifications.
USB Flash Drive
The VALENT Software allows for the transfer of data to a USB external drive.
Warranty
The VALENT warranty is for one year from date of installation and covers all parts and labor when service is performed solely by Biocare
Medical Field Services Personnel. The warranty is invalidated if equipment is abused, damaged, misused, or improperly maintained by
the customer, as determined by Biocare Medical at its sole discretion. The warranty is not transferable to any other party, should the
equipment be resold or transferred by the customer to another party. To the extent permitted by law, Biocare Medical disclaims any
liability for any incidental or consequential damages related to this equipment or for any warranty related services it performs.
Содержание VALENT
Страница 1: ...www biocare net VALENT Manual Automated Slide Staining Platform User Manual ...
Страница 3: ......
Страница 37: ...29 ...
Страница 86: ...78 ...
Страница 115: ...107 ...
Страница 116: ...800 799 9499 60 Berry Drive Pacheco CA 94553 www biocare net VLT8118 010219 V3 ...