Chapter 1: Introduction
Intended Use
This manual is intended for use by clinical, research, industry and reference laboratory personnel who use the VALENT Automated Staining
Platform to perform immunohistochemistry (IHC) staining. This manual describes how to operate the VALENT using the VALENT Software
and will assume users have been trained and have a basic understanding of IHC procedures.
About the VALENT Automated Staining Platform
The VALENT is Biocare Medical’s highest-capacity, fully-automated, open-architecture slide stainer designed to perform high-quality
and consistent IHC staining in lieu of manual staining methods without the variability. However, the VALENT retains the ability to
integrate established manual staining protocols with an automated platform for ease of use with minimal optimization. The VALENT
incorporates on-board deparaffinization and heat-induced antigen retrieval, while expanding capacity to 48 slides and shortening turn-
around-time compared with other auto-stainers.
Limitation of Use
The optimum antibody dilution and protocol for a specific application can vary due to different factors. These may include, but are
not limited to: tissue fixation, antigen retrieval method, incubation times, tissue section thickness and detection kit used. Due to the
superior sensitivity of Biocare Medical’s reagents, the recommended incubation times and titers listed in product data sheets are not
applicable to other detection systems as results may vary and will have to be optimized by the user. The datasheet recommendations
and protocols are based on exclusive use of Biocare Medical products. Ultimately, it is the responsibility of the investigator to determine
optimal staining protocol conditions. The clinical interpretation of any positive or negative staining should be evaluated within the
context of clinical presentation, morphology and other histopathological criteria by a qualified pathologist. The clinical interpretation
of any positive or negative staining should be complemented by morphological studies using proper positive and negative internal and
external controls as well as other diagnostic tests.
VALENT Specifications
Slide capacity
48
Independent slide trays
4 trays, 12 slides per tray
Refrigerated reagent & mixing vial rack
44 positions, 20mL vials;
Temperature range: room temperature to 10°C below room temperature
Antigen retrieval temperature range
Room temperature to 98°C at sea level
Reagent dispensing volume
300µL
Probe dispense capacity
300µL -14.4mL (48 slides x 300µL/slide)
DePar dispense volume per slide
1mL
Bulk reagent carboys
4
Waste carboys
2
Waste separation
Separates hazardous and non-hazardous
Unit control
One instrument per computer
Electrical requirements
Power input: 1800W (1800VA) / Power source: 100-120 VAC, 50/60Hz
The VALENT cannot be used without the VALENT Cart. Power supplied to the VALENT
from the VALENT Cart may reach 3600W@ 120V.
The use of a dedicated circuit capable of delivering a minimum of 1850W continuously
with less than 5% voltage sag is required for the instrument�
Instrument dimensions
40”W X 24”D X 26”H / 102cm W X 61cm D X 66cm H
Instrument weight
245lbs / 111.1 kg