90
Chapter 7: System Cleaning
The VALENT should be cleaned on a daily, weekly, and monthly basis. Failing to clean the instrument may result in poor quality staining.
Daily
• Rinse the used slide racks in tap water and dab (do not wipe) dry with a paper towel after every run. Always wear gloves while
performing this cleaning step.
Weekly
Perform DAB Away cleaning at least once a week or every 250 slides to clean the reagent probe followed by hematoxylin cleaning to
remove residue from troughs and trough level sensors.
Procedure for DAB away cleaning:
Program Cleaning
• From the Home Screen select Tools -> Instrument Operations -> System Cleaning.
• The System Cleaning Window will appear. Follow each instruction and check mark each box for Prerequisites for Cleaning once
the task is complete.