9
Humidity
During Antigen Retrieval humidity levels may increase and create condensation, to prevent this the lab should have an operational HVAC system.
Avoid humidity extremes. Low humidity can cause electrostatic problems that may affect the computer. High humidity may create
condensation.
Waste Disposal
Some of the reagents used on the VALENT may be hazardous, and must be disposed of properly in accordance with local regulations.
The lab should determine from the Safety Data Sheets (SDS) for a given reagent whether the reagent is hazardous. The VALENT can be
programmed to separate waste streams for chromogens considered to be hazardous. The carboy used for this waste stream is marked
“Hazardous Waste.” This carboy must only be used solely for hazardous waste in order to avoid unnecessarily increasing the volume of
any hazardous waste.
In the event a spill of hazardous materials occurs inside the instrument, unplug instrument from the cart and unplug the VALENT Cart
from the wall outlet.
NOTE: All cart circuit breakers must be in the OFF (down) position before disconnecting the power cord from
the cart or the wall outlet� Failure to do so may cause the batteries to run down and necessitate a service call� The user is required to
use PPE which may include a lab coat, gloves, and safety glasses. Use disposable absorbent material to soak up small spills. Follow
approved lab procedures for handling larger spills. Refer to appropriate Safety Data Sheets for proper handling of reagents. The VALENT
Cart has built-in secondary containment for bulk reagent spills with a 11L maximum capacity. If spills occur inside the instrument
contact Biocare Medical’s Technical Support before reconnecting the power.
It is the user's responsibility not to use decontamination or cleaning agents, other than specified in this manual that could cause a
hazard as a result of a reaction with parts of the equipment or with materials used in it. Contact Biocare Medical Technical Support
before using unspecified cleaning agents.
Instrument Disposal
If the VALENT is to be removed from service and disposed of, the instrument must be disinfected using Lysol IC Quaternary Disinfectant
Cleaner through the fluid delivery and waste systems. Similarly, all exposed surfaces must be disinfected using 10% Professional Lysol IC
Quaternary Disinfectant Cleaner before it is disposed of in accordance with local, state and federal regulations.
The VALENT Cart batteries should be re-cycled or properly disposed of in accordance with local regulations.
Site Installation Requirements
Performance checks are conducted by the manufacturer prior to shipping. Biocare Medical Field Service Personnel will unpack or
supervise the unpacking, of the VALENT. Biocare Medical Field Service Personnel will verify the system performance to determine
any deficiencies in operation as a result of shipping and will rectify any problems with regard to performance or components. Biocare
Medical Field Service Personnel will also provide basic operating instructions.
Note: The VALENT weighs approximately 245lbs and must be handled from the bottom of each side of the chassis by two people or
more who are safely capable of lifting the specified weight. Do not lift the instrument from the system lid, doing so will cause damage
to the VALENT and void the warranty.
If the VALENT is placed on a laboratory bench, the bench must be solid, vibration-free with a flat area, capable of safely supporting the
weight of the instrument. Additional consideration must be made for a minimum clearance of 2 inches on all sides for ventilation and
room for an all in one computer, keyboard and mouse next to the instrument. The label and report printers for the VALENT may be located
anywhere that allows them to be connected to the instrument. Biocare Medical will provide a cart for the computer and printers. Nothing
may be placed on top of the VALENT.
The use of a dedicated circuit capable of supplying a minimum of 1850W continuously is required for proper operation. Do not connect
the cart to the same circuit as other equipment, lighting, or any other electrically operated devices.
Support
To contact Biocare Medical Technical Support by telephone, please call 1-800-799-9499, Option 3 or email Technical Support at
[email protected].