Define (Method, Map, Formula and Curve)
|
65
S T A N D A R D S
O N
P L A T E ?
N O
Y
E
S
N
O
If
YES
is chosen, a new standard curve will be generated. The plate map is not
changed. (If “Prompt for Sample ID” is enabled in the UTIL section, you will be
prompted to enter the number of samples. See
Other Utility Options
in
Chapter 3
for more information on the UTIL options.)
If
NO
is chosen, the stored standard curve will be used. If
Auto mapping
had
been used to originally map the standards, blanks, controls and samples
defined for this assay, the map will be automatically regenerated without the
standards, beginning in well xxx (where xxx was chosen as the starting well in
the map, usually well A01). If
manual mapping
was used to map the plate,
the map is
not
regenerated – the reader will
not
produce results for the well
positions that originally were standards. Auto mapping is recommended, if
the standard’s curves will be routinely reused.
Number of Controls
The
NUMBER OF CONTROLS
screen allows you to enter the number of control
groups
that will be used in the assay. Any previously defined value will
appear on the display.
E N T E R
N U M B E R
O F
C
O
N
T
R
O
L
S :
0
2
•
Use the
numeric
keys to enter the
NUMBER OF CONTROLS
groups
in
the assay. For example, if the assay requires one or more positive
control wells and one or more negative control wells, enter 02.
•
The valid entry range depends on the number of locations on the plate
that are undefined. The maximum number of control groups is 8.
ELx800 Operator’s Manual
Содержание ELx800
Страница 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Страница 2: ......
Страница 4: ......
Страница 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Страница 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Страница 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Страница 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Страница 192: ...164 Appendix C Report Format Figure 31 Samples with calls on Matrix Report Bio Tek Instruments Inc...
Страница 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Страница 194: ...166 Appendix C Report Format Figure 33 Samples with calls on Column Report Bio Tek Instruments Inc...
Страница 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Страница 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Страница 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Страница 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Страница 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...