Define (Method, Map, Formula and Curve)
|
55
Mapping Direction
This option allows you to specify how the blank, control, standard, or sample
groups
will be mapped on the plate. The well types can be listed in column
format (down) or in row format (across). The currently selected Mapping
Direction appears on the top line of the display, and the available options
appear on the bottom.
M A P P I N G
D I R E C T I O N : D O W N
D O W N A C R O S S
•
Select
DOWN
to map down the column.
•
Select
ACROSS
to map across the row.
•
Press
ENTER
to save the selection and continue.
Replication Direction
This option allows you to specify how
replicates
are mapped on the plate.
The currently selected Replication Direction appears on the top line of the
display, and the available options appear on the bottom.
R E P
D I R E C T I O N : A C R O S S
D O W N A C R O S S
•
Select
DOWN
to map the replicates down the column, following the
direction of the map listing.
•
Select
ACROSS
to map the replicates across
(in a paired format). As an
example, two replicates can be placed in A1 and A2 wells. The third
replicate would follow in B1. The next standard control, or sample,
would follow in B2.
•
Press
ENTER
to save the selection and continue.
Examples of mapping directions are shown on the next page.
ELx800 Operator’s Manual
Содержание ELx800
Страница 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Страница 2: ......
Страница 4: ......
Страница 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Страница 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Страница 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Страница 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Страница 192: ...164 Appendix C Report Format Figure 31 Samples with calls on Matrix Report Bio Tek Instruments Inc...
Страница 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Страница 194: ...166 Appendix C Report Format Figure 33 Samples with calls on Column Report Bio Tek Instruments Inc...
Страница 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Страница 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Страница 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Страница 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Страница 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...