80
|
Chapter 3: Operation
2.
For a quantitative assay, samples with OD values greater than the
STD2
mean plus a 10% greyzone should be labeled as positive; samples with OD
values less than the
STD2
mean minus the 10% greyzone should be labeled
as negative. All other samples should be considered equivocal.
•
For this example,
STD2;x
(the mean of the STD2 wells) equals 2.000 OD
•
The cutoff formula is simply
STD2;x
•
The greyzone is
10%
•
POS is selected for
SAMP>10%
•
Calls are assigned to sample wells as follows:
EQUIV
if the sample is greater than or equal to 1.800
and
less than or equal to 2.200
POS
if the sample is greater than 2.200
NEG
if the sample is less than 1.800
Bio-Tek Instruments, Inc.
Содержание ELx800
Страница 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Страница 2: ......
Страница 4: ......
Страница 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Страница 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Страница 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Страница 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Страница 192: ...164 Appendix C Report Format Figure 31 Samples with calls on Matrix Report Bio Tek Instruments Inc...
Страница 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Страница 194: ...166 Appendix C Report Format Figure 33 Samples with calls on Column Report Bio Tek Instruments Inc...
Страница 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Страница 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Страница 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Страница 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Страница 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...