144
|
Chapter 6: Troubleshooting and Error Codes
General Errors (Continued)
0400
Carrier X-axis movement failed positional verify
Motor X-axis failed to get to the same position when moved a known number
of steps from the home position and back.
Probable Causes:
•
The belt has slipped because of incorrect tension, a loose motor pulley, or
a loose belt clamp.
•
Defective motor drive.
0401
Carrier Y-axis movement failed positional verify
Motor X-axis failed to get to the same position when moved a known number
of steps from the home position and back.
Probable Causes:
•
The belt has slipped because of incorrect tension, a loose motor pulley, or
a loose belt clamp.
•
Defective motor drive.
0402
Filter wheel failed positional verify; lamp failed
Probable Causes:
•
Optical trigger flag has moved or is loose.
•
Filter wheel is binding against the motor gear.
•
Defective motor drive.
•
Bearings caused motors to fail.
0500 Filter
not
installed
Probable Causes:
•
Empty filter location on the filter wheel (all filter locations must have
either a filter or a filter blank, Bio-Tek PN 2872086).
•
Entire filter wheel is not installed.
Bio-Tek Instruments, Inc.
Содержание ELx800
Страница 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Страница 2: ......
Страница 4: ......
Страница 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Страница 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Страница 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Страница 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Страница 192: ...164 Appendix C Report Format Figure 31 Samples with calls on Matrix Report Bio Tek Instruments Inc...
Страница 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Страница 194: ...166 Appendix C Report Format Figure 33 Samples with calls on Column Report Bio Tek Instruments Inc...
Страница 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Страница 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Страница 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Страница 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Страница 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...