152
|
Appendix A:
Controlling the Reader With KCjunior™ or KC4™
Problems
If KCjunior fails to communicate with the reader, and displays a serial
communications error, check the cable plug-in location to make sure it
matches the setup choices and is not a Null cable. If this is suspected, add
another Null and try again.
Getting Started With KCjunior™
The following instructions
briefly
describe how to read a plate using
KCjunior. Refer to KCjunior's Help system and User's Guide early and
often to learn how to create protocols, assign well identifiers, read plates,
print reports, and more.
To read a plate using KCjunior:
1.
Click
Read Plate
from KCjunior's main screen. The
Read Plate
Dialog will appear.
2.
If desired, enter a Results ID and a Plate Description, and then click
Read Plate.
The
Protocol Definition
dialog will appear.
3.
Select a
Read Method Type
of
Endpoint
,
Kinetic
, or
Multiwavelength.
4.
Define the wavelength(s) at which the plate will be read.
5.
Select a
Plate Geometry
from the drop-down list.
6.
Define other reading parameters as necessary. Click the Help button
for assistance.
7.
When complete, click OK to return to the
Read Plate
dialog. If
desired, enter a
Plate ID.
8.
Place the plate on the carrier, then click
OK
to start the plate read.
The plate will be read and then the raw data results will display
in KCjunior. Print the raw data by selecting Plate|Print Results.
To analyze or manipulate results, a protocol should be defined.
Refer to KCjunior's Help system or User's Guide for instructions.
Bio-Tek Instruments, Inc.
Содержание ELx800
Страница 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Страница 2: ......
Страница 4: ......
Страница 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Страница 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Страница 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Страница 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Страница 192: ...164 Appendix C Report Format Figure 31 Samples with calls on Matrix Report Bio Tek Instruments Inc...
Страница 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Страница 194: ...166 Appendix C Report Format Figure 33 Samples with calls on Column Report Bio Tek Instruments Inc...
Страница 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Страница 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Страница 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Страница 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Страница 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...