Define (Method, Map, Formula and Curve)
|
87
S T A R T
M A P P I N G
A T W E L L L O C A T I O N :
A 0 1
•
Use the
alpha
and
numeric
keys to choose the well location to begin
the assay. Wells A01 through H01 are valid for
ACROSS
mapping; A01
through H12 are valid for
DOWN
.
•
Repeat this process for each assay within the panel. Remain aware of
the total number of controls, standards, and blanks that were originally
mapped in each assay while mapping for the panel assay.
•
For example, to include Assays 1, 8, and 22 in the Panel assay (
DOWN
mapping is selected for the Panel):
Assay 1 has a total of 12 wells defined for controls, blanks, and
standards. In the Panel, the mapping for Assay 1 begins in well A01.
The user wants to run 6 samples in Assay 1. Assay 1 now fills wells
A01 through B03.
The mapping for Assay 8 can begin in well B04, or any well other
than A01 to B03. The reader will “chirp” if you try to map into a
well that is already assigned for use with the Panel.
The mapping for Assay 22 may begin at the next available well
location after Assay 8 mapping is complete.
After all the assays have been entered into the Panel, consider
printing the Panel’s Map Report to verify the map before reading
the plate. Choose
REPORT
(from the Main Menu),
MAP
,
ASSAY 99
.
The reader will print the map of each assay configured in the Panel.
The Panel Assay results are sorted by sample (unless a custom
assay has been programmed by Bio-Tek).
Note:
The interpretation of Results reports for each assay in the Panel
will print first, and then the Sample results will print.
ELx800 Operator’s Manual
Содержание ELx800
Страница 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Страница 2: ......
Страница 4: ......
Страница 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Страница 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Страница 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Страница 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Страница 192: ...164 Appendix C Report Format Figure 31 Samples with calls on Matrix Report Bio Tek Instruments Inc...
Страница 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Страница 194: ...166 Appendix C Report Format Figure 33 Samples with calls on Column Report Bio Tek Instruments Inc...
Страница 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Страница 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Страница 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Страница 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Страница 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...