112
|
Chapter 4: Performance Verification and IQ/OQ/PQ Procedures
Requirements
To run the Absorbance Plate Test, you need Bio-Tek's Absorbance 7-Filter
Test Plate (Part Number 7260522), with its accompanying Data Sheet,
shown in
Figure 18
.
Figure 19
shows a sample Test Plate Analysis
Report.
This test plate can be used for testing the reproducibility, linearity, and
alignment of your Bio-Tek autoreader. The following calibration data has
been recorded by a N.I.S.T. traceable spectrophotometer.
WAVELENGTH (nm)
Well 405nm 450nm 490nm 550nm 620nm 630nm 690nm 750nm
C1 0.147 0.140 0.135 0.130 0.136 0.136 0.127 0.134
E2 0.618 0.575 0.574 0.568 0.573 0.568 0.485 0.434
G3 1.133 1.052 1.051 1.040 1.050 1.040 0.881 0.783
H6 1.701 1.578 1.577 1.560 1.575 1.560 1.323 1.179
F5 2.279 2.024 1.976 1.956 1.893 1.865 1.537 1.272
D4 2.945 2.604 2.545 2.513 2.437 2.400 1.972 1.632
Set # 2453
Serial # 161259
Figure 18: Sample Absorbance Test Plate data sheet
Note:
The Test Plate Analysis report may vary slightly if KC4
or KCjunior
are used.
Absorbance Test Plate Analysis
Wavelength: 405
Alignment Results
B2= 0.000 PASS B12= 0.000 PASS G1= 0.000 PASS G11= 0.000
PASS
Accuracy Results
C01 D04 E02 F05 G03 H06
STANDARD 0.147 2.945 0.618 2.279 1.133 1.701
DATA 0.139 2.914 0.613 2.265 1.128 1.694
RESULT PASS PASS PASS PASS PASS PASS
Repeatability Results
C01 D04 E02 F05 G03 H06
READ1 0.139 2.914 0.613 2.265 1.128 1.694
READ2 0.141 2.913 0.616 2.265 1.128 1.694
RESULT PASS PASS PASS PASS PASS PASS
Figure 19: Sample Test Plate Analysis Report
S A M P L E
Bio-Tek Instruments, Inc.
Содержание ELx800
Страница 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Страница 2: ......
Страница 4: ......
Страница 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Страница 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Страница 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Страница 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Страница 192: ...164 Appendix C Report Format Figure 31 Samples with calls on Matrix Report Bio Tek Instruments Inc...
Страница 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Страница 194: ...166 Appendix C Report Format Figure 33 Samples with calls on Column Report Bio Tek Instruments Inc...
Страница 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Страница 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Страница 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Страница 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Страница 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...