84
|
Chapter 3: Operation
X/Y Axis Type
Note:
LIN/LIN axes are recommended when using the Cubic Spline fit.
After the curve-fit type is selected, select the X/Y Axis Type.
X / Y A X I S
T Y P E :
L I N
L I N
L I N / L O G
L O G
L O G / L I
N
•
Select the method by which the X- and Y-axes will be scaled.
•
This option is not available for the 2-P and 4-P curve-fit types. The X/Y
scaling for these curves is always LIN/LIN.
Extrapolation of Unknowns
This screen allows you to choose whether to extrapolate the curve to evaluate
samples outside of the absorbance range defined by the standards.
E X T R A P O L A T E
U N K N O W N S ? Y E
S
Y E S N O
•
Select
YES
to enable extrapolation; otherwise, select
NO
.
•
On the printed reports, extrapolated concentrations (RSLT values) are
surrounded by
< >
(e.g.,
<44.425>
).
If extrapolation is chosen for the Point-to-Point curve
fit, unknown concentrations will be extrapolated
linearly from the nearest segment of the curve. If the
plot includes both increasing and decreasing segments,
the curve printout will be labeled “Ambiguous.” The
resulting values, which actually are extrapolated, may
not be indicated as such. All calculated results for an
“Ambiguous” curve should be considered unreliable.
Bio-Tek Instruments, Inc.
Содержание ELx800
Страница 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Страница 2: ......
Страница 4: ......
Страница 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Страница 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Страница 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Страница 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Страница 192: ...164 Appendix C Report Format Figure 31 Samples with calls on Matrix Report Bio Tek Instruments Inc...
Страница 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Страница 194: ...166 Appendix C Report Format Figure 33 Samples with calls on Column Report Bio Tek Instruments Inc...
Страница 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Страница 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Страница 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Страница 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Страница 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...