20
|
Chapter 2: Installation
Power-Up and System Test
After you have installed the ELx800 and connected the power supply, turn
on the instrument to run a system test. The on/off switch is located on the
lower right side of the base.
The
System Test
begins with a check of the stepper motors and the analog
power supplies, to ensure that they have a proper input voltage level. The
data flash checksum, motor axis, and analog offset are then verified. The
photodetector’s dark current, noise, and gain are checked to ensure they
fall within specific pass/fail criteria.
If an error is detected, the reader will “chirp” and display an error code.
See
Chapter 6
for a list of error codes. If no errors are detected, the reader
will briefly display
SYSTEM TEST PASS
.
The power-up system test does not produce a printed results report. To run
the test manually and obtain a printout of the system test values, start at
the Main Menu and press
UTIL
TESTS
SYSTEM
.
See
System Test
and
Checksum Test
in
Chapter 4
, or
Connecting a Printer to the
ELx800
on page 22 for more information.
ELx800 Main Menu
Following successful power-up of the ELx800, the
Main Menu
appears:
R E A D Y
0 1 : 2 2 P M
0 1 / 2 3 / 0
4
R E A D
D E F I N E R E P O R T U T I
L
•
The
Main Menu
permits access to all onboard functions. See
Main
Menu
in
Chapter 3
for more information, including a diagram
showing the flow of functionality (
Figure 14
).
•
The ELx800 front panel contains four circular buttons, referred to in
this manual as
“SOFT KEYS.”
One
SOFT KEY
is positioned directly
below each selectable option in the display. To select a menu option,
simply press its corresponding
SOFT KEY
. See
ELx800 Front
Panel
in
Chapter 3
for additional instructions.
Bio-Tek Instruments, Inc.
Содержание ELx800
Страница 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Страница 2: ......
Страница 4: ......
Страница 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Страница 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Страница 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Страница 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Страница 192: ...164 Appendix C Report Format Figure 31 Samples with calls on Matrix Report Bio Tek Instruments Inc...
Страница 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Страница 194: ...166 Appendix C Report Format Figure 33 Samples with calls on Column Report Bio Tek Instruments Inc...
Страница 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Страница 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Страница 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Страница 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Страница 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...