Chapter 4: Performance Verification and IQ/OQ/PQ Procedures
Liquid Testing
Conducting Liquid Tests confirms the ELx800’s ability to perform to
specification with liquid samples. Liquid testing differs from testing with the
Absorbance Test Plate in that liquid in the wells has a meniscus, whereas the
test plate’s neutral density glass filters do not. The optics characteristics may
differ in these two cases, thus alerting the operator to different types of
problems. The liquid tests will help to detect optical defects or contamination
that can contribute to errant readings.
•
Liquid Test 1
confirms repeatability and consistency of the reader
when a solution is used in the microplate. If these tests pass, then the
lens placement and system cleanliness are proven.
•
Liquid Test 2
can be used to test the linearity, repeatability, and
alignment of the reader if you do
not
have an Absorbance Test Plate.
Prepare the series of solutions of varying absorbances as described on
page 121.
•
Liquid Test 3
is an optional test, and can only be performed with the
“UV” model. It is offered for those sites that must have proof of
linearity at wavelengths lower than those attainable with the
Absorbance Test Plate. This test is optional since the ELx800™ has good
“front end” linearity throughout the specified wavelength range (see
page 125).
•
Bio-Tek offers a dye solution (PN 7120779, 25 ml; or 7120782, 125 ml)
that may be used in the stock solution formulation for Liquid Tests 1
and 2, or, if you prefer, you may use the dye solution described in
Table 3
(page 121). The purpose of the formulation is to create a
solution that absorbs light at ~ 2.000 OD full strength when dispensed
at 200
µ
l in a flat-bottom microplate well.
•
Alternatively, any solution that gives a stable color will suffice. (This
includes substrates incubated with an enzyme preparation and then
stopped with an acidic or basic solution.) Some enzyme/substrate
combinations that may be used as alternates to the described dye are
shown in
Table 2.
Table 2
Typical Enzyme-Substrate Combinations and Stopping Solutions
Enzyme
Substrate
Stopping Solution
Alkaline
Phosphate
o-nitrophenyl phosphate
3N sodium hydroxide
beta-
Galactosidase
o-nitrophenyl -beta-D galactopyranoside
1M sodium carbonate
Peroxidase 2,2'-Azino
di-ethylbenzothiazoline-sulfonic
acid (ABTS)
citrate-phosphate
buffer, pH 2.8
Peroxidase
o-phenylenediamine
0.03N sulfuric acid
Bio-Tek Instruments, Inc.
Содержание ELx800
Страница 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Страница 2: ......
Страница 4: ......
Страница 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Страница 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Страница 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Страница 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Страница 192: ...164 Appendix C Report Format Figure 31 Samples with calls on Matrix Report Bio Tek Instruments Inc...
Страница 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Страница 194: ...166 Appendix C Report Format Figure 33 Samples with calls on Column Report Bio Tek Instruments Inc...
Страница 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Страница 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Страница 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Страница 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Страница 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...