Define (Method, Map, Formula and Curve)
|
61
Blank Location
The
BLANK LOCATION
screen allows you to define where the blank well or
wells
occur on the microplate. This screen only appears if
manual mapping
was selected.
E N T E R
T H E
L O C A T I O N
O F
B
L
A
N
K
# 1 :
A
1
2
•
Use the
numeric
and
alpha
keys to enter a Blank Location, based upon
the plate geometry.
•
Use the arrow keys to move the cursor to the next or previous editable
field. The cursor is positioned beneath the first editable field.
•
Press
ENTER
to continue.
Number of Standards
This option allows you to enter the number of standard
groups
that will be
used in the assay. Any previously defined value will be displayed on the
screen.
Note:
If the number of standards is altered, the number of
replicates for the standard automatically reverts
to 1.
E N T E R
N U M B E R
O F
S
T
A
N
D
A
R
D S :
0
2
•
Use the
numeric
keys to enter the
NUMBER OF STANDARDS
. The valid
range depends on the selected curve fit method. The maximum number
of standards is 12. The minimum is 4 for 4-P fit, cubic, cubic spline, and
logit-log; 3 for quadratic; and 2 for linear and point-to-point.
•
Press
CLEAR
to clear the value on the display.
•
Press
ENTER
to continue.
ELx800 Operator’s Manual
Содержание ELx800
Страница 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Страница 2: ......
Страница 4: ......
Страница 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Страница 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Страница 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Страница 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Страница 192: ...164 Appendix C Report Format Figure 31 Samples with calls on Matrix Report Bio Tek Instruments Inc...
Страница 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Страница 194: ...166 Appendix C Report Format Figure 33 Samples with calls on Column Report Bio Tek Instruments Inc...
Страница 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Страница 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Страница 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Страница 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Страница 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...