114
|
Chapter 4: Performance Verification and IQ/OQ/PQ Procedures
To enter the calibration values into KCjunior™:
1.
First define the Absorbance Test Plate to be used for the test. To do
this, open the
Utilities
menu, and select
Diagnostics
|
Universal
Plate Test
.
2.
Click
New Data Sheet
. Click the Help button for guidance.
To enter the calibration values into KC4™:
1.
First define the Absorbance Test Plate to be used for the test. To do
this, open the
System
menu and select
Diagnostics
|
Define
Universal Test Plate|
Add Plate
. Click the Help button for
guidance.
Running the Absorbance Plate Test
IMPORTANT:
Before running the Absorbance Plate
Test, ensure that the reader is not running in
Rapid
Mode. To check this, select
UTIL
READ
and then
cycle through the prompts until
READ IN RAPID
MODE?
is displayed. Choose
NO
for an accurate result.
To run the Absorbance Plate Test from the instrument keypad:
1
Start at the Main Menu and select
UTIL
TESTS
CALPLATE
. The
SELECT TEST
menu will appear:
S E L E C T
T E S T :
S Y S T E M
C H K S U M
C A L P L A T
E
2
Select the appropriate wavelength at the
CALIBRATION FILTER
screen, and press
ENTER
to save the value and continue.
C A L I B R A T I O N
F I L T E R :
4 0 5
4 0 5
4 5 0
4 9 0
* M O R
E
3
When prompted, insert the Test Plate into the plate carrier, and
press the
READ
key to begin the test.
The Calibration Plate Analysis Report (
Figure 19
) will be sent
to a printer when the test is run.
Bio-Tek Instruments, Inc.
Содержание ELx800
Страница 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Страница 2: ......
Страница 4: ......
Страница 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Страница 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Страница 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Страница 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Страница 192: ...164 Appendix C Report Format Figure 31 Samples with calls on Matrix Report Bio Tek Instruments Inc...
Страница 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Страница 194: ...166 Appendix C Report Format Figure 33 Samples with calls on Column Report Bio Tek Instruments Inc...
Страница 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Страница 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Страница 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Страница 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Страница 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...