Controlling the Reader With KCjunior™
|
151
Controlling the Reader With KCjunior™
The ELx800 can be operated using a computer running Bio-Tek's KCjunior
software. Follow the steps below:
1.
Power off the computer and the reader. Connect the serial cable (PN
75053) between the two machines.
2.
Power up both machines.
3.
Install KCjunior on the computer’s hard drive.
4.
Once installed, start KCjunior.
5.
Select
Setup
, then
Reader1.
To select the reader and define the
communications parameters, choose the following setup parameters:
Reader:
ELx800, EL800, or
ELx800UV, EL800UV, or ELx800NB, EL800NB
Com Port:
COM1 or COM2 (the serial port used for the
RS-232 cable connection)
Baud Rate:
2400, 4800, or 9600 must match the baud rate on
the
reader)
Data Bits:
8
Parity: None
Stop Bits:
2
EOT Character: Keep the default number.
6.
Click the
Test Communications
button to attempt to establish
communications with the reader, using the currently defined
communication parameters. If a
Serial Write Error
dialog box is
displayed, an incorrect Com Port may have been selected. Select a
different port and then repeat this step.
7.
If the test passes, click
OK
to save the settings and close the dialog box.
If the test fails, follow the directions provided by KCjunior, then click
Test Communications
again.
To learn more about KCjunior and how to define assays and read plates, refer to
the KCjunior User’s Manual.
ELx800 Operator's Manual
Содержание ELx800
Страница 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Страница 2: ......
Страница 4: ......
Страница 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Страница 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Страница 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Страница 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Страница 192: ...164 Appendix C Report Format Figure 31 Samples with calls on Matrix Report Bio Tek Instruments Inc...
Страница 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Страница 194: ...166 Appendix C Report Format Figure 33 Samples with calls on Column Report Bio Tek Instruments Inc...
Страница 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Страница 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Страница 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Страница 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Страница 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...