Qualification Procedures
|
111
Ensure that the filter calibration values entered are the same as
those on the Test Plate data sheet.
Ensure that the Test Plate is within its calibration certification
period.
If these remedies are ineffective, or if the Test Plate requires
recalibration, call your Bio-Tek representative or Bio-Tek Technical
Support.
•
Repeatability:
This test ensures that the instrument meets its
repeatability specification by reading each Test Plate neutral density
filter twice with the filter in the same location. Note that there may
not be a Pass/Fail indication for filter values that are beyond the
specified accuracy (and, thus, repeatability) range of the device.
If the reader fails the repeatability test, review the following
possible problems and solutions:
Check the filters on the test plate to ensure that there is no debris
that may have shifted between readings and caused changes.
Check the microplate carrier to ensure that it is clear of debris.
If these remedies are ineffective, call your Bio-Tek representative
or Bio-Tek Technical Support.
ELx800 Operator's Manual
Содержание ELx800
Страница 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Страница 2: ......
Страница 4: ......
Страница 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Страница 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Страница 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Страница 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Страница 192: ...164 Appendix C Report Format Figure 31 Samples with calls on Matrix Report Bio Tek Instruments Inc...
Страница 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Страница 194: ...166 Appendix C Report Format Figure 33 Samples with calls on Column Report Bio Tek Instruments Inc...
Страница 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Страница 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Страница 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Страница 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Страница 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...