Liquid Testing
|
117
Stock Solution Formulation
The stock solution for Liquid Tests No. 1 and No. 2 may be formulated
from the ingredients listed below (Solution A), or by diluting a dye solution
available from Bio-Tek (Solution B).
Solution A
Required Materials:
•
Deionized (DI) water
•
FD&C Yellow No. 5 dye powder (typically 90% pure)
•
Tween
®
20 (polyoxyethylene (20) sorbitan monolaurate)
or
Bio-Tek wetting agent, PN 7773002
•
Precision balance with readability of 0.001 g
•
Weigh boat
•
1-liter volumetric flask
Preparation of Stock Solution:
1
Weigh out 0.092 gram of FD&C No. 5 yellow dye powder into a
weigh boat.
2
Rinse the contents into a 1-liter volumetric flask.
3
Add 0.5 ml of Tween 20, or 5 ml of Bio-Tek’s wetting agent.
4
Make up to 1 liter with DI water, cap, and shake well.
This should create a solution with an absorbance of about 2.000
when using 200 µl in a flat-bottom microwell. The OD value will
be proportional to the volume in the well and the amount of
FD&C No. 5 dye used. You can use a larger or smaller well
volume, or add more dye or water to adjust the solution. Note
that too small a well volume may result in increased pipetting-
related errors.
ELx800 Operator's Manual
Содержание ELx800
Страница 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Страница 2: ......
Страница 4: ......
Страница 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Страница 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Страница 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Страница 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Страница 192: ...164 Appendix C Report Format Figure 31 Samples with calls on Matrix Report Bio Tek Instruments Inc...
Страница 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Страница 194: ...166 Appendix C Report Format Figure 33 Samples with calls on Column Report Bio Tek Instruments Inc...
Страница 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Страница 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Страница 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Страница 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Страница 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...