Define (Method, Map, Formula and Curve)
|
59
Blank Map Definitions
AIR
performs an initial reading on “air” just prior to the plate read,
and uses that value as the blank value. This value is subtracted from
each well on the plate.
FULL
enables a single blank well or an average of blank wells to be
subtracted from the whole plate.
CONST
(Constant) allows entry of a user-specified absorbance value.
This value will be subtracted from each well on the plate.
Tip for using
CONST
:
Use a blank value from the first plate, or a
blanking plate, to save space on subsequent assay plates.
ROW
enables a single blank well or an average of blank wells to be
selected for each row. The maximum number of blanks is 48. The
blank (or average) will be subtracted from each well in the row. Use
manual mapping to position blanks, controls, standards, and
samples.
COLUMN
enables a single blank well or an average of blank wells to
be placed in each column. Since the maximum number of blanks is
12, and if all 12 columns are used, each column can have only one
blank. Manual mapping is recommended in this case. Replicates
follow in the same column as the first well of each sample, even if
the ACROSS direction has been specified for replicates.
P-ACROSS
enables a blank in every even-numbered column to be
subtracted from the well to the left of it in every odd column.
Manual mapping is recommended to set up the appropriate map by
placing the standards, controls, and samples in only the odd
columns.
P-DOWN
enables a blank in the B, D, F and H rows to be subtracted
from the well above in the A, C, E and G rows. Manual mapping is
recommended to set up the appropriate map by placing the
standards, controls, and samples in only the A, C, E, and G rows.
ELx800 Operator’s Manual
Содержание ELx800
Страница 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Страница 2: ......
Страница 4: ......
Страница 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Страница 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Страница 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Страница 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Страница 192: ...164 Appendix C Report Format Figure 31 Samples with calls on Matrix Report Bio Tek Instruments Inc...
Страница 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Страница 194: ...166 Appendix C Report Format Figure 33 Samples with calls on Column Report Bio Tek Instruments Inc...
Страница 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Страница 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Страница 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Страница 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Страница 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...