134
|
Chapter 5: Maintenance and Decontamination
To align bulbs on a standard model reader:
Warning!
The lamp will be hot immediately after the
reader has been shut down. Allow the lamp to cool down for
at least 10 minutes before attempting to replace the bulb.
1
Use the lead wires to position the bulb to avoid damaging the glass
covering the bulb with fingertips.
2
Push the bulb forward until it stops.
3
Rotate and/or swing the bulb left or right to get a full circular image
on the carrier sheet metal.
4
Tighten the lamp retention screws and verify that the lamp has not
moved.
5
After completing the alignment, turn the instrument off and on
again to return the carrier to the home position.
Warning!
The alignment procedure
requires you to
observe the light path while the bulb is turned on. To
prevent possible vision impairment, avoid looking directly
at the bulb while it is on.
To align bulbs on UV and NB model readers:
1
Use the lead wires to position the bulb to avoid damaging the glass
covering the bulb with fingertips.
2
Keep the tip of the bulb slightly away from the optics arm #1
aperture (1 mm).
3
Rotate and/or swing the bulb left or right to get a good circular
image on the carrier’s sheet metal.
4
Tighten the lamp retention screws and verify that the lamp has not
moved.
5
After completing the alignment, turn the instrument off and on
again to return the carrier to the home position.
Bio-Tek Instruments, Inc.
Содержание ELx800
Страница 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Страница 2: ......
Страница 4: ......
Страница 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Страница 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Страница 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Страница 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Страница 192: ...164 Appendix C Report Format Figure 31 Samples with calls on Matrix Report Bio Tek Instruments Inc...
Страница 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Страница 194: ...166 Appendix C Report Format Figure 33 Samples with calls on Column Report Bio Tek Instruments Inc...
Страница 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Страница 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Страница 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Страница 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Страница 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...