PN B05577AC
6-8
Sample Modules
Aqueous Liquid Module
Pump Speed
This option opens a dialog box that allows you to set the speed of the pump. Click on the arrows or
drag the indicator to increase or decrease the speed.
Pump On – Pump Off
Selecting one of these options turns the pump ON or OFF.
Open Inlet
Opens the inlet. If the pump is OFF, selecting Open Inlet will turn it ON. If the vessel is already full,
fluid will not be allowed into the vessel.
Close Inlet
Closes the inlet valve regardless of how much fluid is currently in the vessel.
Open Drain
Opens the drain valve. Once the fluid drops below the bottom level sensor, if the pump is ON, the
pump will be turned OFF.
Sonicate
Applies sonication to aid in the dispersion of the sample. Maximum allowable time is 999 seconds.
The maximum output power by the probe is 73 W.
8.1.3.2 Standard Operating Methods (SOM)
The Standard Operating Methods are used in part to assure consistency in the way analyses are
performed. The options available when creating SOMs are similar to those available when using the
Run Cycle.
Creating SOMs
To create an SOM:
Содержание LS 13 320
Страница 4: ...PN B05577AC iv Revision History ...
Страница 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Страница 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Страница 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Страница 58: ...PN B05577AC 2 22 Installation Making Measurements Figure 2 19 Reference Background File Selection ...
Страница 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Страница 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Страница 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Страница 105: ...PN B05577AC 3 47 LS 13 320 Software High Security 3 Figure 3 5 User Privileges Dialog High Security ...
Страница 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Страница 118: ...PN B05577AC 5 2 Data File Menus RunFile Menu 3 Select the file you wish to overlay with the current opened file 4 Open ...
Страница 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Страница 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Страница 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Страница 238: ...PN B05577AC C 10 Optical Models Statistics ...
Страница 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 253: ......