Tornado Dry Powder System
Tornado Description
The Tornado DPS system consists of:
•
The Tornado DPS module
•
A vacuum source
•
A vacuum hose that connects the Tornado DPS to the vacuum
•
Sample tubes
The Tornado DPS provides automatic feed rate (obscuration) control. The set point for the
obscuration is user selectable between 4% and 8%. The accuracy of the average obscuration control
is better than ±2% from the set point. It is measured using the average obscuration set in the Run
Settings dialog box (see the Software section of this manual).
The system disperses cohesive powders without milling fragile materials. The dispersion of the dry
powders is comparable to the dispersion achieved when the samples are run wet with proper
manual pre-dispersion.
The sample to be measured is contained in the system to prevent airborne contamination of the
work area throughout. This is accomplished by maintaining a negative pressure (vacuum) system
for the sample path and trapping the sample via a filtration system in the vacuum.
Indicators
A set of LEDs mounted in the front panel of the module provide visual information about the status
of the system when it is operational. The indicators are active in manual mode as well as indicating
functions accessed through software.
Figure 6.35
Tornado DPS LED Indicators
Error
A red LED which will flash on any error
condition.
Load Sample
Green when sample loading is possible.
Red at all other times while running.
Vacuum
Green when vacuum is within range.
Solid red when vacuum is low.
Flashing red if vacuum is high.
Power
Green when power is applied to the
module's Control PCB.
Содержание LS 13 320
Страница 4: ...PN B05577AC iv Revision History ...
Страница 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Страница 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Страница 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Страница 58: ...PN B05577AC 2 22 Installation Making Measurements Figure 2 19 Reference Background File Selection ...
Страница 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Страница 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Страница 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Страница 105: ...PN B05577AC 3 47 LS 13 320 Software High Security 3 Figure 3 5 User Privileges Dialog High Security ...
Страница 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Страница 118: ...PN B05577AC 5 2 Data File Menus RunFile Menu 3 Select the file you wish to overlay with the current opened file 4 Open ...
Страница 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Страница 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Страница 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Страница 238: ...PN B05577AC C 10 Optical Models Statistics ...
Страница 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 253: ......