PN B05577AC
6-20
Sample Modules
Aqueous Liquid Module
15
Remove the tool and check if the O-ring is completely seated down and around the window.
16
Clean the retaining ring, then gently place it on top of the window.
17
Replace the two new nylon screws that hold the retaining ring.
18
Clean the outer surface of the diffraction window as described in step 10.
19
Replace the window on the other side if needed using steps 4 through 18.
20
Insert both hoses into the bottom of the sample cell stand. Make sure they are tight.
21
Load the ALM sample cell into the optical bench and dock it.
22
Perform a leak test.
a.
Select
Control
,
Fill
.
b.
After the sample vessel is filled, select
Control
,
Pump On
.
c.
If any leaks are noticed after 2 to 3 minutes, remove the retaining ring and check that the
window and O-ring are seated correctly, then repeat the leak test.
Содержание LS 13 320
Страница 4: ...PN B05577AC iv Revision History ...
Страница 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Страница 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Страница 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Страница 58: ...PN B05577AC 2 22 Installation Making Measurements Figure 2 19 Reference Background File Selection ...
Страница 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Страница 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Страница 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Страница 105: ...PN B05577AC 3 47 LS 13 320 Software High Security 3 Figure 3 5 User Privileges Dialog High Security ...
Страница 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Страница 118: ...PN B05577AC 5 2 Data File Menus RunFile Menu 3 Select the file you wish to overlay with the current opened file 4 Open ...
Страница 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Страница 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Страница 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Страница 238: ...PN B05577AC C 10 Optical Models Statistics ...
Страница 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 253: ......