PN B05577AC
E-2
Maintenance
Fourier Lenses
Fourier Lenses
Use this procedure to clean the upper and lower Fourier lenses. To clean the Fourier lenses:
1
If a sample module is docked, undock it, if not, open the Optical Module's door.
CAUTION
Use lens tissues only once, then discard. Never use silicone-coated, eyeglass lens
tissues to clean lenses. They can leave a film. Wash your hands thoroughly before
cleaning any lens or sample cell window. Do not touch optical surfaces with your
fingers or skin. Body oils are difficult to remove from lenses without harming the
anti-reflective coating.
2
Fold three to four tissues in half.
3
Squirt lens cleaner onto the creased area of the folded lens tissue.
4
Place the lens tissues across the upper Fourier lens, holding the tissue in place with two fingers.
Refer to
Figure E.1
Fourier Lens Location and Cleaning Technique
5
Wipe down slowly, keeping the tissues about 3 mm (1/8 in.) below the evaporation line formed.
6
Discard the used lens tissue.
1.
Upper Fourier Lens
2.
Lower Fourier Lens
3.
Tissue
Содержание LS 13 320
Страница 4: ...PN B05577AC iv Revision History ...
Страница 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Страница 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Страница 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Страница 58: ...PN B05577AC 2 22 Installation Making Measurements Figure 2 19 Reference Background File Selection ...
Страница 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Страница 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Страница 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Страница 105: ...PN B05577AC 3 47 LS 13 320 Software High Security 3 Figure 3 5 User Privileges Dialog High Security ...
Страница 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Страница 118: ...PN B05577AC 5 2 Data File Menus RunFile Menu 3 Select the file you wish to overlay with the current opened file 4 Open ...
Страница 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Страница 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Страница 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Страница 238: ...PN B05577AC C 10 Optical Models Statistics ...
Страница 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 253: ......