PN B05577AC
2-13
Installation
Making Measurements
2
Running a Control
After performing the installation functions, the instrument is ready to measure the light intensity
patterns and calculate the particle size distribution. The above steps are recommended when
analyzing a sample in order to obtain consistent and accurate results. The selection of the controls
(provided by Beckman Coulter) to be used will depend on the sample module being used and are
listed in
. The following section describes the procedure to follow in order to verify the
operation of the instrument by running a control sample.
1
Run
.
2
Run Cycle
(as shown in
Figure 2.9
Run Menu
Table 2.3
Controls Used With Specific Sample Modules
Sample Module
Control
Universal Liquid Module (ULM)
Latron 300LS, G15, GB 500
Aqueous Liquid Module (ALM)
Latron 300LS, G15, GB 500
Tornado Dry Powder
G35, GB 500
Micro Liquid Module
G15
Содержание LS 13 320
Страница 4: ...PN B05577AC iv Revision History ...
Страница 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Страница 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Страница 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Страница 58: ...PN B05577AC 2 22 Installation Making Measurements Figure 2 19 Reference Background File Selection ...
Страница 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Страница 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Страница 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Страница 105: ...PN B05577AC 3 47 LS 13 320 Software High Security 3 Figure 3 5 User Privileges Dialog High Security ...
Страница 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Страница 118: ...PN B05577AC 5 2 Data File Menus RunFile Menu 3 Select the file you wish to overlay with the current opened file 4 Open ...
Страница 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Страница 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Страница 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Страница 238: ...PN B05577AC C 10 Optical Models Statistics ...
Страница 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 253: ......