Sample Modules
Universal Liquid Module
6
6
In Step 4 the run length is set. For analyses that require collection of PIDS data, a run length of
at least 90 seconds is recommended. For analyses without PIDS, 60 seconds is the recommended
time for the analysis, though analyses may be as short as 10 seconds. The optical model can be
selected during this step. For more on optical models see
. The Pump Speed can be
adjusted in this dialog. It is recommended for samples that require surfactant that the pump
speed be kept low, i.e. 30%, in order to prevent the formation of air bubbles. Should bubbles be
present, alternating high pump speed with low pump speed is useful when trying to de-bubble
the suspending fluid. If the pump speed is increased to its maximum (100%) and then reduced
to about 50%, air bubbles can be eliminated. This procedure may need to be repeated two to
three times to eliminate the bubbles. High pump speeds are also useful when measuring large,
dense particles. You can access the manual speed control from the Control menu by selecting
Pump Speed.
Содержание LS 13 320
Страница 4: ...PN B05577AC iv Revision History ...
Страница 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Страница 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Страница 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Страница 58: ...PN B05577AC 2 22 Installation Making Measurements Figure 2 19 Reference Background File Selection ...
Страница 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Страница 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Страница 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Страница 105: ...PN B05577AC 3 47 LS 13 320 Software High Security 3 Figure 3 5 User Privileges Dialog High Security ...
Страница 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Страница 118: ...PN B05577AC 5 2 Data File Menus RunFile Menu 3 Select the file you wish to overlay with the current opened file 4 Open ...
Страница 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Страница 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Страница 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Страница 238: ...PN B05577AC C 10 Optical Models Statistics ...
Страница 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 253: ......