PN B05577AC
B-4
Sample Handling
Liquid Sample Dispersion
Scoop sampling
This technique is prone to error since the whole sample does not pass through the sampling device.
Sample is taken from the surface, where it may not be typical of the bulk. To eliminate segregation,
it is usual to shake the sample in a container using different modes of shaking.
Figure B.4
Figure 4. Scoop Sampling
Cone and quartering
Cone and quartering consists of pouring the material into a conical heap and hoping that all four
parts are symmetrical. If symmetry does not exist in the cutting and flattening of the heap, size
error will occur. This method is greatly dependent on the skill of the operator and consequently
should not be used.
Figure B.5
Figure 5. Cone and Quartering
Liquid Sample Dispersion
Dispersion can make or break the analysis regardless of the particle analyzer being used in the
experiment. It allows us to see the correct particle distribution for a given sample. Factors that
govern the dispersion method that is to be utilized are:
•
Solubility (e.g., fertilizer)
•
Reactivity (e.g., cement, lithium hydride with water)
•
Suspendibility (e.g., use of glycerol)
•
End use (e.g., pigments, flour)
Содержание LS 13 320
Страница 4: ...PN B05577AC iv Revision History ...
Страница 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Страница 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Страница 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Страница 58: ...PN B05577AC 2 22 Installation Making Measurements Figure 2 19 Reference Background File Selection ...
Страница 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Страница 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Страница 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Страница 105: ...PN B05577AC 3 47 LS 13 320 Software High Security 3 Figure 3 5 User Privileges Dialog High Security ...
Страница 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Страница 118: ...PN B05577AC 5 2 Data File Menus RunFile Menu 3 Select the file you wish to overlay with the current opened file 4 Open ...
Страница 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Страница 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Страница 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Страница 238: ...PN B05577AC C 10 Optical Models Statistics ...
Страница 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 253: ......