PN B05577AC
2-5
Installation
Instrument Verification
2
Instrument Verification
Post-Installation Verification
To verify the performance of your LS 13 320, a Beckman Coulter control sample must be analyzed.
The acceptable limits are listed on the assay sheets enclosed with the control samples.
See
Daily Verification
Beckman Coulter recommends that you run a control sample daily to verify your instrument's
performance. Calibration is determined by the optical design. Therefore, no calibration is required.
Measuring electrical offsets and aligning the laser beam makes all necessary adjustments.
See
Beckman Coulter offers a series of control samples to check the operation of your instrument. A
quality control program is also available, which is a convenient way to have the performance of your
instrument tracked and any problems identified by qualified Beckman Coulter personnel.
Getting Started
The PC software has been designed to control all the operations of the LS 13 320. Once the
instrument and its corresponding module have been set up and the software installed, click on the
program icon to launch the program. When starting up the instrument after it has been shut down,
allow the instrument to warm up for approximately 20 minutes.
Starting the Instrument and its Control Program
When the program is first launched, the first view presented by the program is the copyright
acknowledgement box (
OK
launches the control program.
shows the status screen after selecting
OK
from
the above screen.
Содержание LS 13 320
Страница 4: ...PN B05577AC iv Revision History ...
Страница 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Страница 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Страница 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Страница 58: ...PN B05577AC 2 22 Installation Making Measurements Figure 2 19 Reference Background File Selection ...
Страница 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Страница 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Страница 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Страница 105: ...PN B05577AC 3 47 LS 13 320 Software High Security 3 Figure 3 5 User Privileges Dialog High Security ...
Страница 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Страница 118: ...PN B05577AC 5 2 Data File Menus RunFile Menu 3 Select the file you wish to overlay with the current opened file 4 Open ...
Страница 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Страница 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Страница 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Страница 238: ...PN B05577AC C 10 Optical Models Statistics ...
Страница 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 253: ......