PN B05577AC
6-59
Sample Modules
Tornado Dry Powder System
6
Cleaning the Cell Windows
To clean the windows the sample cell must be removed from the module housing.
1
Loosen the screws that hold the bracket in place. Lift the bracket to disengage the pins from the
pin slots. See
.
Figure 6.36
Bracket Screws
2
Undo the four (top and bottom) latches found in the back of the module that attach the sample
cell to the module housing (
1.
Pin
2.
Bracket
3.
Slot
4.
Screw Securing the Bracket
Содержание LS 13 320
Страница 4: ...PN B05577AC iv Revision History ...
Страница 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Страница 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Страница 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Страница 58: ...PN B05577AC 2 22 Installation Making Measurements Figure 2 19 Reference Background File Selection ...
Страница 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Страница 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Страница 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Страница 105: ...PN B05577AC 3 47 LS 13 320 Software High Security 3 Figure 3 5 User Privileges Dialog High Security ...
Страница 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Страница 118: ...PN B05577AC 5 2 Data File Menus RunFile Menu 3 Select the file you wish to overlay with the current opened file 4 Open ...
Страница 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Страница 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Страница 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Страница 238: ...PN B05577AC C 10 Optical Models Statistics ...
Страница 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 253: ......