Installation
Hardware Connections
2
Drain for Aqueous Liquid Module
Connect the drain hose to a wastewater drain located near the instrument.
Drain for Universal Liquid Module
If a suspension fluid other than water is used, connect the drain hose to a dedicated waste disposal
container. It is your responsibility to properly dispose of any hazardous samples and suspension
fluids used in the instrument.
Vacuum for Tornado Dry Powder System
The Dry Powder System requires a vacuum source (optionally supplied by Beckman Coulter) for
operation. If you choose to supply one, you will need a unit capable of sustaining rates up to
420 L/minute at 740 to 750 torr (15 cfm at 5 to 10 inches of water below atmospheric pressure). A
remote power switch and hoses are also needed to connect your vacuum source to the Dry Powder
System. Your Beckman Coulter Representative can recommend a vacuum source for your system.
Hardware Connections
for the location of interunit connections as well as the on/off switch on the Optical
Bench. The location of ports on the computer, monitor and printer may vary according to the
hardware you purchase with your unit. Check the appropriate manufacturer's manual.
Figure 2.2
Location of Interunit Connections
1.
Computor Interface Cable
2.
ALM, DPS and Autoprep Station
Содержание LS 13 320
Страница 4: ...PN B05577AC iv Revision History ...
Страница 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Страница 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Страница 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Страница 58: ...PN B05577AC 2 22 Installation Making Measurements Figure 2 19 Reference Background File Selection ...
Страница 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Страница 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Страница 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Страница 105: ...PN B05577AC 3 47 LS 13 320 Software High Security 3 Figure 3 5 User Privileges Dialog High Security ...
Страница 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Страница 118: ...PN B05577AC 5 2 Data File Menus RunFile Menu 3 Select the file you wish to overlay with the current opened file 4 Open ...
Страница 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Страница 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Страница 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Страница 238: ...PN B05577AC C 10 Optical Models Statistics ...
Страница 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 253: ......