PN B05577AC
6-49
Sample Modules
Tornado Dry Powder System
6
Figure 6.34
Figure 17 Cleaning the Cell With a Swab
MLM Troubleshooting
Tornado Dry Powder System
Problem
Possible Cause/Solution
No communication with optical bench
•
Module docked incorrectly. Eject module and re-dock.
Stirrer will not turn
•
Liquid too viscous
•
Cell not mounted properly
•
O-rings missing from stirrer bar
Sensor light not on
•
Module not docked properly. Eject module and re-dock.
•
Cell not mounted properly
Obscuration too high
•
Dirty windows. Clean windows.
Module leaking
•
Check fittings
The Tornado Dry Powder System is intended for use with the
LS 13 320 Optical Bench. It is capable of feeding and measuring
dry powder samples in the size range 0.4 µm to 2000 µm. The
Tornado DPS measures the entire sample presented to the
instrument, with a sample volume that allows an analysis that
will provide statistically accurate results. The range of sample
volumes the instrument can accept should provide for a
minimum 10 second run at controlled obscuration. There is no
need for sample preparation and operator intervention is
minimal. The system is controlled via software commands that
make it completely hands-free. The sample is placed in a sample
holder and delivered to the sensing zone in the optical bench by
a vacuum.
Содержание LS 13 320
Страница 4: ...PN B05577AC iv Revision History ...
Страница 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Страница 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Страница 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Страница 58: ...PN B05577AC 2 22 Installation Making Measurements Figure 2 19 Reference Background File Selection ...
Страница 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Страница 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Страница 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Страница 105: ...PN B05577AC 3 47 LS 13 320 Software High Security 3 Figure 3 5 User Privileges Dialog High Security ...
Страница 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Страница 118: ...PN B05577AC 5 2 Data File Menus RunFile Menu 3 Select the file you wish to overlay with the current opened file 4 Open ...
Страница 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Страница 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Страница 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Страница 238: ...PN B05577AC C 10 Optical Models Statistics ...
Страница 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 253: ......