PN B05577AC
B-9
Sample Handling
Diluent Selection
B
Consider Characteristics of the Diluent
Viscosity:
1.
Too high:
•
Entraining of bubbles
•
Inadequate sampling of particles due to slow movement of diluent
•
Pump may work too hard
•
Difficult to disperse sample in diluent properly
2.
Too low may cause formation of explosive mixture of air and vapor
Active life:
1.
Will the diluent absorb unwanted water?
2.
Will the diluent allow growth of bacteria, yeast, etc. (may use preservatives)?
Reactivity:
1.
Will the diluent degrade the instrument's polyethylene fill and waste tanks over time?
2.
Will the diluent degrade the instrument's stainless steel fill and waste tanks over time?
3.
Will the diluent degrade any wetted materials in the instrument?
Purity:
1.
Is the diluent chemically pure?
2.
Is the diluent free of interfering particles? (If not, filtering may be appropriate.)
Refractive Index:
Is the R.I. of the diluent the same or very close to that of the sample? If so, the instrument may not
be able to see the sample.
Color:
1.
If the color of the diluent is similar to the color of the laser, the instrument may not be able to
see the sample.
2.
If the color of the diluent causes the sample to be too opaque, the instrument may not be able
to see the sample.
Содержание LS 13 320
Страница 4: ...PN B05577AC iv Revision History ...
Страница 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Страница 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Страница 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Страница 58: ...PN B05577AC 2 22 Installation Making Measurements Figure 2 19 Reference Background File Selection ...
Страница 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Страница 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Страница 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Страница 105: ...PN B05577AC 3 47 LS 13 320 Software High Security 3 Figure 3 5 User Privileges Dialog High Security ...
Страница 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Страница 118: ...PN B05577AC 5 2 Data File Menus RunFile Menu 3 Select the file you wish to overlay with the current opened file 4 Open ...
Страница 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Страница 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Страница 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Страница 238: ...PN B05577AC C 10 Optical Models Statistics ...
Страница 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 253: ......