PN B05577AC
6-15
Sample Modules
Aqueous Liquid Module
6
Cleaning the ALM
It is important for all glass surfaces in your sample system to be clean. Dusty or coated optical
surfaces can cause erroneous results. Some samples or suspension fluids may coat the inside of the
sample cell windows. If the instrument is in a dusty or smoky environment, or if vapors are present,
the lenses and outer surfaces of the windows may become coated. Check the optical surfaces for
cleanliness at least every three months, and clean if needed. Depending on the environment
cleaning may be required more frequently.
Cleaning Fluids
Beckman Coulter recommends cleaning solution supplied with the module for cleaning the inner
surfaces of the ALM sample system. Beckman Coulter recommends lens cleaning paper for cleaning
the lenses and sample cell windows.
Cleaning the Cell
Clean the inner surfaces of the sample system (pump, hoses and within the sample cells) whenever
they become coated.
To clean the sample system:
1
Select
Control
.
2
Select
Rinse
. Rinse for approximately five minutes with water. Select
Cancel
to complete the
rinsing routine.
3
Select
Control
again.
4
Select
Open Drain
.
5
Select
Close Drain
.
6
Prepare the cleaning solution.
7
Fill the vessel with cleaning solution.
8
Let the cleaning solution circulate for 15 minutes to 1 hour, depending on how dirty the cell is.
9
Select
Open Drain > Rinse
.
10
The windows can be cleaned by removing the thumbscrews that hold the cover shown in
Содержание LS 13 320
Страница 4: ...PN B05577AC iv Revision History ...
Страница 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Страница 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Страница 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Страница 58: ...PN B05577AC 2 22 Installation Making Measurements Figure 2 19 Reference Background File Selection ...
Страница 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Страница 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Страница 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Страница 105: ...PN B05577AC 3 47 LS 13 320 Software High Security 3 Figure 3 5 User Privileges Dialog High Security ...
Страница 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Страница 118: ...PN B05577AC 5 2 Data File Menus RunFile Menu 3 Select the file you wish to overlay with the current opened file 4 Open ...
Страница 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Страница 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Страница 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Страница 238: ...PN B05577AC C 10 Optical Models Statistics ...
Страница 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 253: ......