PN B05577AC
6-46
Sample Modules
Micro Liquid Module
4
To remove air bubbles after the sample cell has been filled, tilt the cell to the side to allow
trapped air to escape. Tilt it the opposite direction and repeat this procedure. Alternatively,
place the cell on the sample stand and turn the stirring motor on at a low speed. With high
viscosity fluids, you may have to turn off the motor periodically to let the bubbles rise.
5
Screw the cap securely onto the cell. Ensure that the O-ring is in place before screwing on the
cap.
6
Place the cell back on the MLM stand.
Adding Sample
Sample is added to the sample cell until the correct obscuration is obtained. To add sample:
1
Open the Optical Bench door.
2
Remove the cell cap and add a small amount of sample to the sample cell.
3
Close the door and note the obscuration.
4
Repeat the previous two steps until the correct obscuration is obtained.
IMPORTANT
Do not allow liquid to spill inside the optical bench while loading sample into the cell.
Setting the Stirring Motor
The speed of the stirring motor may need to be adjusted, depending on the sample being analyzed.
For most applications, maximum speed is appropriate. Lower speeds may allow larger or denser
particles to settle, skewing the size distribution. Lower speeds have been provided so that delicate
samples, or those likely to flocculate when agitated, can be run.
Содержание LS 13 320
Страница 4: ...PN B05577AC iv Revision History ...
Страница 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Страница 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Страница 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Страница 58: ...PN B05577AC 2 22 Installation Making Measurements Figure 2 19 Reference Background File Selection ...
Страница 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Страница 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Страница 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Страница 105: ...PN B05577AC 3 47 LS 13 320 Software High Security 3 Figure 3 5 User Privileges Dialog High Security ...
Страница 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Страница 118: ...PN B05577AC 5 2 Data File Menus RunFile Menu 3 Select the file you wish to overlay with the current opened file 4 Open ...
Страница 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Страница 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Страница 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Страница 238: ...PN B05577AC C 10 Optical Models Statistics ...
Страница 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 253: ......