PN B05577AC
xxviii
LS 13 320 Introduction
System Components
Figure 6
Mechanical Spatial Filter
Sample Modules
The main function of the sample-handling module is to deliver particles in the sample, without
discrimination to their sizes, to the sensing zone while avoiding the introduction of any undesirable
effects such as air bubbles and/or thermal turbulence. The sample module is typically composed of
the sample cell and a delivery system. The delivery system may include certain features such as a
circulation pump, an ultrasonic probe, or stirring bars to help better disperse and circulate the
particles. The LS 13 320 operates with sample cells designed for particles suspended in liquids or in
dry powder form. These modules are:
•
Tornado Dry Powder System (DPS)
•
Universal Liquid Module (ULM)
•
Aqueous Liquid Module (ALM)
•
Micro Liquid Module (MLM)
In addition the ALM can be used in conjunction with the AutoPrep Station.
Computer System
The operation of the LS 13 320 requires a personal computer (PC) and the PC-based control and
analysis software. If you are not using a Beckman Coulter-supplied PC, you must provide one that
meets the minimum configuration requirements shown in
, and preferably one that meets
the recommended configuration requirements (shown in the same table).
1.
Input Beam Profile
2.
Scattering from Dust
Particles
3.
Objective
4.
“Dirty” Spatial Profile
5.
Pin Hole
6.
“Clean” Spatial Profile
Table 1
Minimum and Recommended Computer Configurations
Item
Minimum Configuration
Recommended Configuration
Microprocessor
800 MHz
1 GHZ Intel or equivalent
RAM
512 Mb
1 Gb or better
Содержание LS 13 320
Страница 4: ...PN B05577AC iv Revision History ...
Страница 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Страница 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Страница 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Страница 58: ...PN B05577AC 2 22 Installation Making Measurements Figure 2 19 Reference Background File Selection ...
Страница 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Страница 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Страница 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Страница 105: ...PN B05577AC 3 47 LS 13 320 Software High Security 3 Figure 3 5 User Privileges Dialog High Security ...
Страница 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Страница 118: ...PN B05577AC 5 2 Data File Menus RunFile Menu 3 Select the file you wish to overlay with the current opened file 4 Open ...
Страница 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Страница 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Страница 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Страница 238: ...PN B05577AC C 10 Optical Models Statistics ...
Страница 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 253: ......