PN B05577AC
6-58
Sample Modules
Tornado Dry Powder System
Loading an SOP
To load an SOP:
1
Run
.
2
Select Load an SOP. SOPs can also be loaded from the status panel by clicking on the
Load SOP
button.
3
Select the desired SOP file from the SOP subfolder or the subfolder where the file was last saved.
Cleaning the Tornado DPS
It is important for all glass surfaces in your sample system to be clean, as dusty or coated optical
surfaces can cause erroneous results. Some samples may coat the inside of the sample cell windows.
If the instrument is in a dusty or smoky environment or if vapors are present, the lenses and outer
surfaces of the windows may become coated. Check the optical surfaces for cleanliness at least every
three months, and clean if needed. Depending on the environment, cleaning may be required more
frequently.
Using the Cleaning Sample
The sample cell and tubing may be cleaned by using the provided cleaning sample. Load
approximately 15 cc to 25 cc of the cleaning sample into the sample holder and select the
Run
Cleaner
option from the Control menu.
Содержание LS 13 320
Страница 4: ...PN B05577AC iv Revision History ...
Страница 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Страница 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Страница 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Страница 58: ...PN B05577AC 2 22 Installation Making Measurements Figure 2 19 Reference Background File Selection ...
Страница 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Страница 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Страница 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Страница 105: ...PN B05577AC 3 47 LS 13 320 Software High Security 3 Figure 3 5 User Privileges Dialog High Security ...
Страница 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Страница 118: ...PN B05577AC 5 2 Data File Menus RunFile Menu 3 Select the file you wish to overlay with the current opened file 4 Open ...
Страница 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Страница 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Страница 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Страница 238: ...PN B05577AC C 10 Optical Models Statistics ...
Страница 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 253: ......