PN B05529AE
3-5
Troubleshooting
Emergency Access
3
4
Set the covers aside.
5
Insert a Phillips-head screwdriver through one of the exposed holes in the centrifuge front
panel and loosen, but do not remove, the screw.
a.
Repeat for the other screw.
6
Move the centrifuge forward on the bench until the two screws on the bottom of the centrifuge
are accessible (see
Figure 3.1
).
7
Use a Phillips-head screwdriver to remove the two screws and set them aside.
8
Carefully lift the centrifuge front cover until the bottom releases, then pull the top towards you
slightly.
To support the cover, insert the cover bottom lip under the standoff plate inside the centrifuge.
CAUTION
Do not allow the cover to hang down supported only by the exposed power and
grounding wires or the wires will be damaged.
9
Pull the latch-release cable firmly to the right until the latches release and the door opens
(see
Figure 3.2
).
Figure 3.2
Internal Access
10
If the rotor is still spinning, close the door and wait until it stops before attempting to remove it.
1.
Latch release cable.
2.
Standoff plate.
3.
Cover bottom lip goes here.
Содержание Allegra X-30 Series
Страница 4: ...PN B05529AE iv Revision History ...
Страница 8: ...PN B05529AE viii Safety Notice Chemical and Biological Safety ...
Страница 14: ...xiv Tables ...
Страница 18: ...PN B05529AE xviii Introduction CE Mark ...
Страница 30: ...PN B05529AE 1 12 Description Available Rotors ...
Страница 56: ...PN B05529AE 4 6 Care and Maintenance Supply List ...
Страница 62: ...PN B05529AE Warranty 2 Beckman Coulter Inc Allegra X 30 Series Centrifuge Warranty ...
Страница 63: ......