PN B05529AE
1-10
Description
Available Rotors
Available Rotors
See the applicable rotor manual for information on rotor use, care and maintenance, and rotor
accessories.
Electrical supply
Class I
Maximum heat dissipation into room under
steady-state conditions
3311 Btu/h (0.97 kW)
Noise level 0.91 m (3 ft) in front of
instrument (approx.)
£68 dBa
Installation (overvoltage) category
II
Pollution degree
2
b
a. Temperature range depends on rotor in use and speed (see applicable rotor manual).
b. Normally only nonconductive pollution occurs; occasionally, however, a temporary conductivity caused by condensation must be expected.
Rotor Profile
Description
Refrigerated
Nonrefrigerated
Max
Capacity
(mL)
Rotor Part
Number/
Rotor Manual
Number
Max
RPM
Max RCF
(
g
)
Max
RPM
Max
RCF
(
g
)
F2402H
a
Fixed Angle
45° Angle
r
max
= 82 mm
18,000
29,756
16,000
23,511
24
1.5/2.0
361171
GS-TB-021
FX301.5
Fixed Angle
45° Angle
r
max
= 100 mm
16,000
b
28,672
b
13,200
19,515
30
1.5/2.0
392274
MMR-TB-002
F0630
Fixed Angle
30° Angle
r
max
= 78 mm
18,000
28,305
16,000
22,364
6
30
361231
GS-TB-014
F0850
Fixed Angle
25° Angle
r
max
= 94 mm
11,400
13,682
9000
8528
8
50
346640
GS-TB-003
F0685
Fixed Angle
25° Angle
r
max
= 97 mm
10,000
10,864
8000
6953
6
85
364650
GS-TB-008
Содержание Allegra X-30 Series
Страница 4: ...PN B05529AE iv Revision History ...
Страница 8: ...PN B05529AE viii Safety Notice Chemical and Biological Safety ...
Страница 14: ...xiv Tables ...
Страница 18: ...PN B05529AE xviii Introduction CE Mark ...
Страница 30: ...PN B05529AE 1 12 Description Available Rotors ...
Страница 56: ...PN B05529AE 4 6 Care and Maintenance Supply List ...
Страница 62: ...PN B05529AE Warranty 2 Beckman Coulter Inc Allegra X 30 Series Centrifuge Warranty ...
Страница 63: ......