PN B05529AE
4-3
Care and Maintenance
Tube Breakage
4
Tube Breakage
If a glass tube breaks, and all the glass is not contained in the bucket or rotor, you must thoroughly
clean the interior of the chamber bowl.
WARNING
Be careful when examining or cleaning the sealing gasket or chamber, as sharp
glass fragments may be embedded in their surfaces.
1
Examine the gasket to make sure that no glass particles are retained in it.
a.
Carefully remove any glass particles that may remain.
2
Carefully wipe away any glass particles that remain in the bowl.
Decontamination
If the centrifuge and/or accessories are contaminated with radioactive or pathogenic solutions,
perform appropriate decontamination procedures. Refer to
Chemical Resistances
(IN-175) to be sure
the decontamination method will not damage any part of the centrifuge.
Sterilization and Disinfection
The centrifuge is finished with urethane paint. Ethanol (70%)
*
may be used on this surface. See
Chemical Resistances
for more information regarding chemical resistance of centrifuge and
accessory materials.
While Beckman Coulter has tested these methods and found that they do not damage the
centrifuge, no guarantee of sterility or disinfection is expressed or implied. When sterilization or
disinfection is a concern, consult your laboratory safety officer regarding proper methods to use.
* Flammability hazard. Do not use in or near operating centrifuges.
Содержание Allegra X-30 Series
Страница 4: ...PN B05529AE iv Revision History ...
Страница 8: ...PN B05529AE viii Safety Notice Chemical and Biological Safety ...
Страница 14: ...xiv Tables ...
Страница 18: ...PN B05529AE xviii Introduction CE Mark ...
Страница 30: ...PN B05529AE 1 12 Description Available Rotors ...
Страница 56: ...PN B05529AE 4 6 Care and Maintenance Supply List ...
Страница 62: ...PN B05529AE Warranty 2 Beckman Coulter Inc Allegra X 30 Series Centrifuge Warranty ...
Страница 63: ......